Analog Input and Output Calibration Procedures
C-2
4. Connect the UDAS to the host computer and allow at least a 15 minute warm-up period before
performing the adjustments.
FIGURE C.1 Locations of Potentiometers for Adjusting Analog Circuits on page C-2 shows the
locations of the potentiometers used to adjust the analog circuits on a UDAS unit.
FIGURE C.1 Locations of Potentiometers for Adjusting Analog Circuits
CAUTION:
The unit may become damaged if you use an uninsulated screwdriver
and accidentally touch other components.
00 00 0 0
H4 H4
R24
R25
R26
TP21
R39
R38
R40
R46
R45
R36
TP27
R58